RSAP 180 Learning Outcome #8 Review
Authored by Kaylyn Grummett
Science, Physics
University
Used 34+ times

AI Actions
Add similar questions
Adjust reading levels
Convert to real-world scenario
Translate activity
More...
Content View
Student View
20 questions
Show all answers
1.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
The most critical factor in obtaining diagnostic quality images using automatic exposure control (AEC) is the use of accurate____________.
Positioning and centering
Focal spot size
Side markers
Backup time
Answer explanation
Automatic exposure devices provide diagnostic quality exposures only for structures positioned directly above the ionization chambers. If the anatomy isn’t properly placed directly above the ionization chamber, the resulting image can be under or over exposed.
The art of using AEC is the art of positioning
2.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
TRUE or FALSE: the preset amount of radiation exposure is reached sooner with high kVp.
True
False
Answer explanation
Because high kVp radiation is more penetrating (reducing the total amount of x-ray exposure to the patient because more x-ray photons exit the patient) and the detectors are measuring quantity of radiation, the preset amount of radiation exposure is reached sooner with high kVp.
3.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
The ionization detectors that are used in most automatic exposure controlled units are located between _________ and ___________.
Tube; collimator
Collimator; patient
Filter; grid
Patient; image receptor
Answer explanation
Modern AECs are ionization chambers, located before the image receptor and after the grid.
4.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
TRUE OR FALSE: When using AEC, the radiographer loses control over kVp.
True
False
Answer explanation
When utilizing AEC, the radiographer loses control over time, and as a result mAs when using an AEC.
5.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
If you have too much collimation and collimate an ionization chamber out, the resulting image would be:
Creates an under exposed image
Creates an over exposed image
No affect
System will not allow an exposure
Answer explanation
If the primary beam is collimated from an active chamber, the chamber operates as if the tissue is extremely dense (dark image)
6.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
TRUE OR FALSE: AEC is a system used to consistently control the amount of radiation reaching the image receptor by terminating the length of exposure
True
False
7.
MULTIPLE CHOICE QUESTION
45 sec • 1 pt
The purpose of the backup timer in an AEC system is to:
Prevent excessive exposure by shutting the exposure off in case the AEC's timer switch failure
Controls the length of time required for the AEC chamber to respond to the radiation and terminate the exposure
Determines the total amount of mAs required to produce the appropriate exposure to the image receptor
Permits the choice of an anatomical region and the projection
Answer explanation
Minimum Response Time: Controls the length of time required for the AEC chamber to respond to the radiation and terminate the exposure
mAs Readout: Determines the total amount of mAs required to produce the appropriate exposure to the image receptor
APR: Permits the choice of an anatomical region and the projection