1.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
How to validate that imprecision for both IQC and human samples is similar?
Compared variance using F-Test
Do Lot-to-lot reagent verification
Partition the IQC that have shift or significant change
Assuming by eye ball
2.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
Partition of IQC data need to calculate separately due to
QC lot that cause a significant shift in QC value without change in patient sample
Insufficient data for long term interval of IQC
Change of calibrator lot
MU not fit the purpose
3.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
What is formula of MU estimation when bias is not include in calculation
u = √ (uRw2 + ucal2 + uBias2)
u = √ (uRw2 + ucal2)
u = √ (uRw2 )
All of above
4.
MULTIPLE SELECT QUESTION
45 sec • 1 pt
Uncertainty due to bias correction, ubias is rarely include in MU calculation due to (more than 1 answer)
Limitation of bias calculation using EQA bias
The laboratory applying correction factor
Small changes in bias will be captures by long term precision
All above
5.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
Reporting of MU more consistent across concentration intervals using this
Absolute uncertainty (u)
Long term precision, URW
Relative uncertainty , Urel
Calibrator uncertainty, Ucal