1.

MULTIPLE CHOICE QUESTION

30 sec • 1 pt

How to validate that imprecision for both IQC and human samples is similar?

Compared variance using F-Test

Do Lot-to-lot reagent verification

Partition the IQC that have shift or significant change

Assuming by eye ball

2.

MULTIPLE CHOICE QUESTION

30 sec • 1 pt

Partition of IQC data need to calculate separately due to

QC lot that cause a significant shift in QC value without change in patient sample

Insufficient data for long term interval of IQC

Change of calibrator lot

MU not fit the purpose

3.

MULTIPLE CHOICE QUESTION

30 sec • 1 pt

What is formula of MU estimation when bias is not include in calculation

u = √ (uRw2 + ucal2 + uBias2)

u = √ (uRw2 + ucal2)

u = √ (uRw2 )

All of above

4.

MULTIPLE SELECT QUESTION

45 sec • 1 pt

Uncertainty due to bias correction, ubias is rarely include in MU calculation due to (more than 1 answer)

Limitation of bias calculation using EQA bias

The laboratory applying correction factor

Small changes in bias will be captures by long term precision

All above

5.

MULTIPLE CHOICE QUESTION

30 sec • 1 pt

Reporting of MU more consistent across concentration intervals using this

Absolute uncertainty (u)

Long term precision, URW

Relative uncertainty , Urel

Calibrator uncertainty, Ucal

Access all questions and much more by creating a free account

Create resources

Host any resource

Get auto-graded reports

Google

Continue with Google

Email

Continue with Email

Classlink

Continue with Classlink

Clever

Continue with Clever

or continue with

Microsoft

Microsoft

Apple

Apple

Others

Others

Already have an account?

Discover more resources for Professional Development