Advance MU Part 2

Advance MU Part 2

1st - 3rd Grade

5 Qs

quiz-placeholder

Similar activities

Grade2 Islamic 3R

Grade2 Islamic 3R

2nd Grade

10 Qs

LGBTQ

LGBTQ

KG - Professional Development

9 Qs

Prep-Cast CAPM Domain#1 PM Fundamental & Core Concepts

Prep-Cast CAPM Domain#1 PM Fundamental & Core Concepts

1st Grade

10 Qs

Bao Đồng 2

Bao Đồng 2

1st Grade

8 Qs

SAN ZHI XIAO ZHU -xiaozhu he fangzi

SAN ZHI XIAO ZHU -xiaozhu he fangzi

KG - 1st Grade

10 Qs

PUPR

PUPR

1st - 3rd Grade

10 Qs

Ch-2 Under Ideal  Accounting-Scott

Ch-2 Under Ideal Accounting-Scott

KG - Professional Development

10 Qs

DMAIC - Measure Tools

DMAIC - Measure Tools

1st - 10th Grade

10 Qs

Advance MU Part 2

Advance MU Part 2

Assessment

Quiz

Professional Development

1st - 3rd Grade

Medium

Created by

Noor Yacob

Used 4+ times

FREE Resource

5 questions

Show all answers

1.

MULTIPLE CHOICE QUESTION

30 sec • 1 pt

How to validate that imprecision for both IQC and human samples is similar?

Compared variance using F-Test

Do Lot-to-lot reagent verification

Partition the IQC that have shift or significant change

Assuming by eye ball

2.

MULTIPLE CHOICE QUESTION

30 sec • 1 pt

Partition of IQC data need to calculate separately due to

QC lot that cause a significant shift in QC value without change in patient sample

Insufficient data for long term interval of IQC

Change of calibrator lot

MU not fit the purpose

3.

MULTIPLE CHOICE QUESTION

30 sec • 1 pt

What is formula of MU estimation when bias is not include in calculation

u = √ (uRw2 + ucal2 + uBias2)

u = √ (uRw2 + ucal2)

u = √ (uRw2 )

All of above

4.

MULTIPLE SELECT QUESTION

45 sec • 1 pt

Uncertainty due to bias correction, ubias is rarely include in MU calculation due to (more than 1 answer)

Limitation of bias calculation using EQA bias

The laboratory applying correction factor

Small changes in bias will be captures by long term precision

All above

5.

MULTIPLE CHOICE QUESTION

30 sec • 1 pt

Reporting of MU more consistent across concentration intervals using this

Absolute uncertainty (u)

Long term precision, URW

Relative uncertainty , Urel

Calibrator uncertainty, Ucal